The D-RC 120 is an automatic sampling device for gravimetric dust measurements to adjust or calibrate dust monitors for the measurement of dust in mg/m3.
The D-RC 120 extracts isokinetically a partial gas stream from the measuring gas duct whereas the extraction is controlled fully automatically. The partial gas stream is sucked through a filter in which the contained dust is deposited. The dust concentration is determined by weighing the filter before and after the extraction.
The design of the D-RC 120 meets the requirements of DIN EN 13284-1 and US-EPA Method 5.
Краткий обзор
- Portable system
- Easy handling
- Automatic isokinetic sampling
- Suitable for in-stack and out-stack sampling
- Highest measuring accuracy
- Probe available in stainless steel and titanium
- Compliant with EN 13284-1, VDI 2066, EN ISO 9096 and US-EPA Method 5
примене́ние
- Calibration of dust monitors in mg/m3 according to DIN EN 13284-1 and VDI 2066
- Adjustment of dust monitors before performing QAL 2
- Determination of dust concentration at stacks for internal quality control
- Measurements at filter plants
Технический данные
Sample gas type | Air, flue gas, non-flammable process gas |
sample gas temperature | 0 ... 500 °C |
Sample gas dust concentration | 0.1 ... 2000 mg/m3 |
Inner duct pressure | -200 ... +20 hPa relative to ambient |
Sample gas relative humidity | < 95% rel. humidity, noncondensing, optional measurement in saturated flue gas below the dew point |
Sample gas flow velocity | 2 ... 39 m/s |
Duct inner diameter | 0.35 ... 6 m |
Measuring principle | Gravimetric analysis |
Conformities | EN 13284-1, VDI 2066, EN ISO 9096, US-EPA Method 5 |
Measurement | Dust concentration |
Measuring location | In-stack, optional out-stack |
System components | D-RC 120 MD Automatic isokinetic sampling device - D-RC 120 control unit - Pump - Pulsation damper - Dryer tower - External sensor unit - Windows software, accessories D-RC 120 FD filter head device, suitable for plane filters or filter thimbles D-RC 120 P combination probe, heatable, incl. gooseneck with nozzles |
Physical measuring values | Dust concentration Differential pressure Absolute pressure Temperature Partial volume flow |
Measuring range dust concentration | 0.1 ... 2000 mg/m3 |
Measuring range differential pressure | 0.01 ... 10 hPa |
Measuring range absolute pressure | 400 ... 1.300 hPa |
Measuring range temperature | 0 ... 500 °C |
Measuring range partial volume flow | 0.5 ... 4 m3/h |
Electrical interfaces | 1x 12 V output for piezo siren 1x 4...20 mA input 1x RS 485 with USB converter for PC communication |
Mechanical interfaces | |
Power supply | 230 VAC, 50/60 Hz, 115 VAC, 50/60 Hz as an option |
Operation | Windows software, requires netbook/laptop with Windows 7 or 8 |
Ambient temperature (operational) | -10 ... +60°C |
IP protection class (IEC 60529) | IP 33 |
Electrical safety | CE |
Dimensions (h x w x d) | Approx. 350 x 280 x 250 mm |
Weight | Approx. 8.5 kg |
Type | Plane Filter Head Device, Tubular-Plane Filter Head Device as alternative |
Dimensions Plane Filter Head Device (lxd) | 120 mm x 55 mm |
Dimensions Tubular-Plane Filter Head Device (lxd) | 145 mm x 55 mm |
Nozzle diameter | 6, 8, 10, 12, 14 mm |
Diameter plane filter | 45 mm |
Length filter thimble | 68 mm |
Weight | Approx. 1 kg without gooseneck and nozzle |
Material | Stainless steel, optional titanium |
Probe length | 1000 ... 3000 mm, others on request |
Probe diameter | 70 mm, quick relase clamp for 3" port |
Weight | 1.5 kg + 5.4 kg/m |
Material | Stainless steel, optional titanium |
Subject to technical modifications!